A dual-wavelength spectrometer uses 780 nm and 830 nm. The molar extinction coefficients for oxy- hemoglobin (HbO2) and deoxy-hemoglobin (Hb) at these two wavelengths are: e_HbO2_780nm = 710 cm 1M-1, e_Hb_780nm = 1075 cm-1M-1, e_HbO2_830nm = 974 cm-1M-1, e_Hb_830nm = 693 cm-1M-1. The spectrometer measures absorption coefficients at these two wavelengths are: Ua_780nm = 0.057 cm-", Ua_830nm = 0.063 cm-, please calculate the concentration for HbO2 and Hb (in unit of mM).

a. [HbO2] = 18.45 uM; [Hb] = 12.11 UM

b. [HbO2] = 8.45 UM; [Hb] = 22.11 UM

c. [Hb] = 8.45 uM; [HbO2] = 22.11 UM

d. [Hb] = 19.44 UM; [HbO2] = 50.85 UM

answer
Answers: 1

Other questions on the subject: Engineering

image
Engineering, 03.07.2019 14:10, volleyballfun24
If the thermal strain developed in polyimide film during deposition is given as 0.0044. assume room temperature is kept at 17.3 c, and thermal coefficient of expansion for the film and the substrate are 54 x 10^-6c^-1 and 3.3 x 10^-6c^-1respectively. calculate the deposition temperature.
Answers: 3
image
Engineering, 04.07.2019 18:10, redrosesxx
Water at 55c flows across a flat plate whose surface temperature is held constant at 95c. if the temperature gradient at the plate's surface for a given value of x is 18 c/mm, find a) local heat transfer coefficient. b) heat flux
Answers: 3
image
Engineering, 04.07.2019 18:10, salazjlove
Which of the following refers to refers to how well the control system responds to sudden changes in the system. a)-transient regulation b)- distributed regulation c)-constant regulation d)-steady-state regulation
Answers: 1
image
Engineering, 04.07.2019 18:20, esmemaluma00
Find the minimum film thickness for a journal bearing with the data below. shaft diameter, d-50 mm, clearance ratio, cdratio? 0.001, shaft speed, n 2000 rpm; bearing length. i 200 mm; eccentricity ration, ? -0.55. ( note, cdratio-ca/d) the minimum film thickness is um
Answers: 2
Do you know the correct answer?
A dual-wavelength spectrometer uses 780 nm and 830 nm. The molar extinction coefficients for oxy- he...

Questions in other subjects:

Konu
English, 28.02.2021 21:10